ACQUISITION SETTINGS
====================
The released frames are cropped to the acoustic sector, so the machine's
on-screen annotation of these settings is not visible in them. The values below
were read from that annotation across all 720 acquisitions before it was
removed. Fields that varied between acquisitions are recorded per file in
acquisitions.csv.
System and transducer
---------------------
Ultrasound system Esaote MyLab A70
Transducer C 1-8A curvilinear
Preset Abdominal, General
Mode B-mode, with tissue enhancement imaging (TEI) engaged
Native export 1200 x 932 pixels
Cine capture rate 31 frames per second, approximately 5 seconds per loop
Constant across all 720 acquisitions
------------------------------------
Imaging depth 166 mm (displayed by the machine as "D 166 mm")
Acoustic power 100%
General map Gen-M
Persistence / filter PRC 9/1/1/0
Transmit / multiline X/M +2/4
Pulse repetition PRS 6
Frame rate readout 32 Hz
Note on depth: the machine reports 166 mm. Where a nominal depth of 15 cm has
been quoted for this study elsewhere, 166 mm is the value the system recorded.
Varying between sessions
------------------------
Three fields were not constant. Per-file values are given in acquisitions.csv
as gain_mode, mechanical_index and thermal_index_soft_tissue.
Gain mode automatic (AG) 354 of 720 acquisitions
preset (G) 366 of 720
Mechanical index 1.3 315
1.4 240
0.9 103
1.1 62
Thermal index (soft 0.3 480
tissue) 0.2 240
Gain was set once per session and not adjusted during scanning: the machine's
automatic gain was engaged for some participants and the preset gain left in
place for others. It is a per-participant property, not a per-image one.
Fourteen participants were scanned entirely with automatic gain and fifteen
entirely with the preset. The single exception is ogxjyv, where the fasted state
was acquired on the preset and automatic gain was engaged from the clear-fluid
state onward.
Gain mode is balanced across the four gastric states (47-50% automatic in each)
and across the two body positions (49% in both), and is statistically
independent of state (chi-square 0.6 on 3 degrees of freedom). It therefore
carries no information about what the stomach contained.
The mechanical and thermal indices are readouts that follow from the gain and
focus in use rather than settings chosen directly.